120 - 344 Litre Front Loading Autoclave
Large front loading autoclave
Model: ASB270 Heaters in chamber (153 litres)
Model: ASB280 Heaters in chamber (247 litres)
Model: ASB290 Heaters in chamber (290 litres)
Model: ASB300 Heaters in chamber (344 litres)
120 - 344 Litre Front Loading Autoclave
The 120 - 344 litre front loading autoclave offers five chamber sizes, increased in capacity from the top loading range. The 'Swiftlock' front loading range of autoclaves allows the user a greater choice of options unavailable on the smaller ranges, as well as easy loading access.
Astell’s additional options include Integral Steam Generators, Heated Jackets, Simple/Advanced Vacuum and a choice of Cooling options to name but a few. Learn more about each of these by viewing the pages below or clicking the View Product Options button:
Astell autoclaves are used in scientific research worldwide. View selected peer-reviewed research papers that reference their use in experimental methodology.
Contact us now on +44 (0)20 8309 2031 to find out more, or request a quotation by clicking here. For organisations outside the UK, please contact your national dealer who can be found on our Find a Dealer page.
Range Classification
A standard Swiftlock unit is fitted with heaters in the base of the chamber and is ideal for sterilizing liquids (media), discard, glassware and other instruments. Advanced options such as the Integral Steam Generator, External Jacket and choice of Vacuum can be applied to customise a vessel for many other applications such as liquids, fabrics and other porous loads.
Direct steam models are available at no extra cost.
Astell can custom build the Swiftlock range to your specifications.
This product has received Sustainability Awards from SLS. Click to find out more.
View Product Options Download Brochure View Products in Stock
Features
- Recirculating coolant conservation system with temperature-controlled tundish
- 316 stainless steel electropolished chamber
- A choice of 120, 153, 247, 290 and 344 litres models
- Calibration and Validation available
- FDA 21 CFR part 11 software option available
- Cooling Lock
- External pressure guage
- Five password levels to stop lower level users making changes
- Holdwarm / Delayed Start feature
- Optimised for low water and power consumption
- Over Temperature Protection
- Standard Astell safety features
- Timed/Pulsed Free Steaming
- Fully programmable 5.7" Colour Touch Screen controller
- USB interface for downloading controller data
Product Options
Product Downloads
SLS Sustainable Features Awards
This autoclave has been recognized by Scientific Laboratory Supplies for its outstanding sustainable features through the following awards.
Sustainable Raw Material Award
Material resource efficiency principles have been embedded into the design.
Sustainable Product Manufacturing Award
Feature of manufacturing process designed to generate less waste.
Sustainable Distribution & Packaging Award
Manufacturing facility located in UK or Ireland. Country of origin has an Environmental Performance Index (EPI) greater than 60 (2022). Plastic free packaging. Product reduces need for packaging. Packaging return scheme offered.
Sustainable Use Award
Product feature reduces energy consumption. Product feature reduces water consumption. Built to last. Minimum 5+ year warranty with software support (if applicable) and spare parts and servicing available.
Sustainable Disposal Award
Product is recyclable (supplier must offer a recycling or repurposing scheme for lab products) – excludes packaging.
Icons courtesy of Scientific Laboratory Supplies
Specifications and Options for 120–344 Litre Front Loading Autoclaves
Astell autoclaves can be equipped with a wide range of optional components, enabling them to meet your exact specifications without unnecessary hardware. Units can also be custom-built to suit the unique demands of specific sites and applications. Below are the standard specifications for each model listed.
ASB260 - 120 Litre Front Loading Autoclave
Specifications
Air (7)
| Specification | Value | Unit |
|---|---|---|
| Connecting Airline Hose Size For Air Ballast | TBC | mm |
| Peak Air Flow For Air Ballast | 100 | Litres per minute |
| Peak Air Pressure For Air Ballast | TBC | bar |
| Second Air Supply Use | simple or advanced vacuum | |
| Second Airline Supply Connection Size | TBC | mm |
| Second Peak Air Supply Flow | TBC | Litres per minute |
| Second Peak Air Supply Pressure | TBC | bar |
Chamber (6)
| Specification | Value | Unit |
|---|---|---|
| Chamber Depth | 740 | mm |
| Chamber Finish | Electropolished | |
| Chamber Height | 454 | mm |
| Chamber Volume | 120 | Litres |
| Chamber Width | 454 | mm |
| Load Height | 790 | mm |
Dimensions (5)
| Specification | Value | Unit |
|---|---|---|
| Dry Weight | 125.5 | kg |
| Product Depth | 1150 | mm |
| Product Footprint | 0.79 | square metres |
| Product Height | 1265 | mm |
| Product Width | 686 | mm |
Doors (1)
| Specification | Value | Unit |
|---|---|---|
| Number Of Doors | 1 |
Drain (6)
| Specification | Value | Unit |
|---|---|---|
| Drain Material | Copper | |
| Drain Position | above floor | |
| Drain Size | 35 | mm |
| Drainpipe Connection | Between Autoclave and Main Building Drain Only | |
| Maximum Drain Height | 690 | mm |
| Required Features of Drainage System | Venting |
Electrical (6)
| Specification | Value | Unit |
|---|---|---|
| Max Frequency | 60 | Hz |
| Max Voltage | 230 | V |
| Max Wattage | 18 | kW |
| Power Of Steam Source | TBC | kW |
| Standard Amperage | 30 | A |
| Three Phase Availability | Available |
Process (2)
| Specification | Value | Unit |
|---|---|---|
| Fitted Vacuum System | Optional | |
| Steam Source | In-chamber water heating element |
Water (3)
| Specification | Value | Unit |
|---|---|---|
| Maximum Water Flow Rate | Infinity | Litres per minute |
| Minimum Water Flow Rate | 0 | Litres per minute |
| Water Enters Via | a water jug, filled from a tap |
Options
Options (32)
| Code | Name | Description | Functionality / Features |
|---|---|---|---|
| AAQ503EXT | External Air Compressor | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| DSS | Steam Separator | Enhances the quality of steam supplied to the autoclave, ensuring its suitability for sterilization. | |
| RSSC70 | Secondary Controller | A secondary screen used on double-door pass-through autoclaves, allowing the operator to view the controller displayed information from the opposite door. |
|
| 16KWSG | 16kW Integral Steam Generator | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| AVC001S | Advanced Vacuum | Provides pre- and post-sterilization vacuum phases, using negative and positive pressure pulsing to remove air from the chamber and load before sterilization, and a drying stage after sterilization. When combined with a Chamber Vessel Heating Jacket, drying performance is enhanced. |
|
| AVC004 | Air Ballast | Maintains chamber pressure during the cooling of fluid loads, greatly reducing the likelihood of boil-over. Requires suitable compressed air, and works best when paired with Load-Sensed Process Timing and a Cooling Jacket. |
|
| AVC007 | Special Pathogen Free Seal | A mechanical seal used in double-door (pass-through) autoclaves, uniting the autoclave with a surrounding wall to maintain separation between the clean and unclean sides of a laboratory. Prevents cross-contamination, allowing the autoclave to function as a pathogen-controlled access point between two locations. |
|
| AVQ006S | CAT III (CL3 / BSL3) Compliance | Enhanced containment methods prevent unsterilized pathogens from leaving the autoclave, meeting the requirements of CAT III compliance for Containment Level 3 / Biosafety Level 3. |
|
| CFR021 | FDA 21 CFR Part 11 Compliant Controller Software | Provides compliance to the requirements of 21 CFR Part 11 through data security, data integrity, traceability/audit trails, and electronic signature rules. |
|
| AAN420 | Drain Cooling | Cools effluent (waste water) before it leaves the autoclave. For sites without thermal-shock-resistant drains. Requires a mains water supply. |
|
| AAP006 | Chamber Vessel Cooling Fan System | Blows air across the external surfaces of the sterilization chamber vessel to improve cooling times. |
|
| AAP018 | Autodrain | Automatically drains the in-chamber steam generation assembly at the end of a cycle, allowing the sterilization chamber to cool faster. |
|
| AAP019 | Autofill | Allows automatic refilling of the in-chamber steam generation assembly. Requires mains water supply. |
|
| AAP100 | Chamber Vessel Cooling Water Coils | Pipework coiled around the sterilization chamber vessel allows water to be circulated around the outside, improving cooling times. |
|
| AAP102S | In Chamber Fan | Circulates gases inside the chamber during cooling, allowing the load to cool extremely quickly. Requires a Chamber Vessel Cooling Jacket for optimal performance, and Air Ballast is required when sterilizing liquids. Also required for over-pressure (Steam-Air mix) applications. |
|
| AAQ503 | Integral Air Compressor | Provides compressed air for autoclave systems that require it. | |
| AAR014 | Load Sensed Process Timing | A temperature probe with an armoured lead is added to the chamber. Placed inside fluid loads, this probe provides accurate measurement of liquid temperatures, allowing for faster and more controlled liquid sterilization. |
|
| AAR122 | Ethernet Interface | Allows the autoclave controller to be monitored from compatible digital systems. |
|
| AAR130 | Integral Data-Printer | Allows for the generation of physical records. |
|
| AJP100 | Chamber Vessel Heating Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with steam to heat the chamber walls. Provides improved heating speeds, and load drying after sterilization. |
|
| AJP100C | Chamber Vessel Cooling Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with water to cool the chamber walls. Provides rapid cooling. |
|
| AVC002S | Simple Vacuum | Provides a vacuum phase before and after sterilization to aid with air removal and load drying, reducing time to sterilization. Suitable for use with in-chamber steam generation. |
|
| AAQ525 | External Air Compressor | Provides compressed air for autoclave systems that require it. |
|
| CWS300 | Water Softener and Installation Kit | Extracts limescale-causing minerals from the water supply before they enter the autoclave, reducing the likelihood of limescale fouling the autoclave. |
|
| AAB001 | Blow Down Vessel | For use with an integral or stand-alone steam generator. The autoclave can be set to expel high-pressure steam to remove dissolved gases and debris from its components; this vessel safely captures the expelled condensate and debris during the blow-down process. |
|
| AAN080 | Container Tray | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN300 | Morrison Discard Container | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN316 | Additional Shelf | An additional shelf on anti-tilt brackets, typically placed in the centre of the autoclave, providing approximately twice the usable width and depth of chamber space. |
|
| UKAS001 | UKAS Calibration | Calibration of the autoclave to specified temperatures as per the requirements of ISO 17025:2005. |
|
| IQ/OQ001 | IQ/OQ Documents Package | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| IQ/OQ003 | Completion of IQ/OQ Documentation | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| UKASVAL | Load Validation | Multipoint Load Validation Service provided by an Astell Services Autoclave Engineer. |
|
ASB270 - 153 Litre Front Loading Autoclave
Specifications
Air (7)
| Specification | Value | Unit |
|---|---|---|
| Connecting Airline Hose Size For Air Ballast | TBC | mm |
| Peak Air Flow For Air Ballast | 100 | Litres per minute |
| Peak Air Pressure For Air Ballast | TBC | bar |
| Second Air Supply Use | simple or advanced vacuum | |
| Second Airline Supply Connection Size | TBC | mm |
| Second Peak Air Supply Flow | TBC | Litres per minute |
| Second Peak Air Supply Pressure | TBC | bar |
Chamber (6)
| Specification | Value | Unit |
|---|---|---|
| Chamber Depth | 945 | mm |
| Chamber Finish | Electropolished | |
| Chamber Height | 454 | mm |
| Chamber Volume | 153 | Litres |
| Chamber Width | 454 | mm |
| Load Height | 790 | mm |
Dimensions (5)
| Specification | Value | Unit |
|---|---|---|
| Dry Weight | 231.5 | kg |
| Product Depth | 1350 | mm |
| Product Footprint | 0.93 | square metres |
| Product Height | 1265 | mm |
| Product Width | 686 | mm |
Doors (1)
| Specification | Value | Unit |
|---|---|---|
| Number Of Doors | 1 |
Drain (6)
| Specification | Value | Unit |
|---|---|---|
| Drain Material | Copper | |
| Drain Position | above floor | |
| Drain Size | 35 | mm |
| Drainpipe Connection | Between Autoclave and Main Building Drain Only | |
| Maximum Drain Height | 690 | mm |
| Required Features of Drainage System | Venting |
Electrical (7)
| Specification | Value | Unit |
|---|---|---|
| Max Frequency | 60 | Hz |
| Max Voltage | 230 | V |
| Max Wattage | 10 | kW |
| Power Of Steam Source | TBC | kW |
| Single Phase Availability | Available | |
| Standard Amperage | 40 | A |
| Three Phase Availability | Available |
Process (2)
| Specification | Value | Unit |
|---|---|---|
| Fitted Vacuum System | Optional | |
| Steam Source | In-chamber water heating element |
Water (3)
| Specification | Value | Unit |
|---|---|---|
| Maximum Water Flow Rate | Infinity | Litres per minute |
| Minimum Water Flow Rate | 0 | Litres per minute |
| Water Enters Via | a water jug, filled from a tap |
Options
Options (32)
| Code | Name | Description | Functionality / Features |
|---|---|---|---|
| AAQ503EXT | External Air Compressor | The standard screw down legs of the autoclave are replaced with castors allowing for easy movement and positioning of the autoclave. |
|
| DSS | Steam Separator | Enhances the quality of steam supplied to the autoclave, ensuring its suitability for sterilization. | |
| RSSC70 | Secondary Controller | A secondary screen used on double-door pass-through autoclaves, allowing the operator to view the controller displayed information from the opposite door. |
|
| AVC001S | Advanced Vacuum | Provides pre- and post-sterilization vacuum phases, using negative and positive pressure pulsing to remove air from the chamber and load before sterilization, and a drying stage after sterilization. When combined with a Chamber Vessel Heating Jacket, drying performance is enhanced. |
|
| AVC002S | Simple Vacuum | Provides a vacuum phase before and after sterilization to aid with air removal and load drying, reducing time to sterilization. Suitable for use with in-chamber steam generation. |
|
| AVC004 | Air Ballast | Maintains chamber pressure during the cooling of fluid loads, greatly reducing the likelihood of boil-over. Requires suitable compressed air, and works best when paired with Load-Sensed Process Timing and a Cooling Jacket. |
|
| AVC007 | Special Pathogen Free Seal | A mechanical seal used in double-door (pass-through) autoclaves, uniting the autoclave with a surrounding wall to maintain separation between the clean and unclean sides of a laboratory. Prevents cross-contamination, allowing the autoclave to function as a pathogen-controlled access point between two locations. |
|
| AVQ006S | CAT III (CL3 / BSL3) Compliance | Enhanced containment methods prevent unsterilized pathogens from leaving the autoclave, meeting the requirements of CAT III compliance for Containment Level 3 / Biosafety Level 3. |
|
| CFR021 | FDA 21 CFR Part 11 Compliant Controller Software | Provides compliance to the requirements of 21 CFR Part 11 through data security, data integrity, traceability/audit trails, and electronic signature rules. |
|
| AAN420 | Drain Cooling | Cools effluent (waste water) before it leaves the autoclave. For sites without thermal-shock-resistant drains. Requires a mains water supply. |
|
| AAP006 | Chamber Vessel Cooling Fan System | Blows air across the external surfaces of the sterilization chamber vessel to improve cooling times. |
|
| AAP018 | Autodrain | Automatically drains the in-chamber steam generation assembly at the end of a cycle, allowing the sterilization chamber to cool faster. |
|
| AAP019 | Autofill | Allows automatic refilling of the in-chamber steam generation assembly. Requires mains water supply. |
|
| AAP100 | Chamber Vessel Cooling Water Coils | Pipework coiled around the sterilization chamber vessel allows water to be circulated around the outside, improving cooling times. |
|
| AAP102S | In Chamber Fan | Circulates gases inside the chamber during cooling, allowing the load to cool extremely quickly. Requires a Chamber Vessel Cooling Jacket for optimal performance, and Air Ballast is required when sterilizing liquids. Also required for over-pressure (Steam-Air mix) applications. |
|
| AAQ503 | Integral Air Compressor | Provides compressed air for autoclave systems that require it. | |
| AAR014 | Load Sensed Process Timing | A temperature probe with an armoured lead is added to the chamber. Placed inside fluid loads, this probe provides accurate measurement of liquid temperatures, allowing for faster and more controlled liquid sterilization. |
|
| AAR122 | Ethernet Interface | Allows the autoclave controller to be monitored from compatible digital systems. |
|
| AAR130 | Integral Data-Printer | Allows for the generation of physical records. |
|
| AJP100 | Chamber Vessel Heating Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with steam to heat the chamber walls. Provides improved heating speeds, and load drying after sterilization. |
|
| AJP100C | Chamber Vessel Cooling Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with water to cool the chamber walls. Provides rapid cooling. |
|
| 16KWSG | 16kW Integral Steam Generator | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| AAB001 | Blow Down Vessel | For use with an integral or stand-alone steam generator. The autoclave can be set to expel high-pressure steam to remove dissolved gases and debris from its components; this vessel safely captures the expelled condensate and debris during the blow-down process. |
|
| AAQ525 | External Air Compressor | Provides compressed air for autoclave systems that require it. |
|
| CWS300 | Water Softener and Installation Kit | Extracts limescale-causing minerals from the water supply before they enter the autoclave, reducing the likelihood of limescale fouling the autoclave. |
|
| AAN300 | Morrison Discard Container | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN316 | Additional Shelf | An additional shelf on anti-tilt brackets, typically placed in the centre of the autoclave, providing approximately twice the usable width and depth of chamber space. |
|
| AAN080 | Container Tray | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| IQ/OQ001 | IQ/OQ Documents Package | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| IQ/OQ003 | Completion of IQ/OQ Documentation | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| UKAS001 | UKAS Calibration | Calibration of the autoclave to specified temperatures as per the requirements of ISO 17025:2005. |
|
| UKASVAL | Load Validation | Multipoint Load Validation Service provided by an Astell Services Autoclave Engineer. |
|
ASB280 - 247 Litre Front Loading Autoclave
Specifications
Air (7)
| Specification | Value | Unit |
|---|---|---|
| Connecting Airline Hose Size For Air Ballast | TBC | mm |
| Peak Air Flow For Air Ballast | 100 | Litres per minute |
| Peak Air Pressure For Air Ballast | TBC | bar |
| Second Air Supply Use | simple or advanced vacuum | |
| Second Airline Supply Connection Size | TBC | mm |
| Second Peak Air Supply Flow | TBC | Litres per minute |
| Second Peak Air Supply Pressure | TBC | bar |
Chamber (5)
| Specification | Value | Unit |
|---|---|---|
| Chamber Depth | 876 | mm |
| Chamber Finish | Electropolished | |
| Chamber Height | 600 | mm |
| Chamber Volume | 247 | Litres |
| Chamber Width | 600 | mm |
Dimensions (5)
| Specification | Value | Unit |
|---|---|---|
| Dry Weight | 387 | kg |
| Product Depth | 1440 | mm |
| Product Footprint | 1.3 | square metres |
| Product Height | 1420 | mm |
| Product Width | 902 | mm |
Doors (1)
| Specification | Value | Unit |
|---|---|---|
| Number Of Doors | 1 |
Drain (6)
| Specification | Value | Unit |
|---|---|---|
| Drain Material | Copper | |
| Drain Position | above floor | |
| Drain Size | 35 | mm |
| Drainpipe Connection | Between Autoclave and Main Building Drain Only | |
| Maximum Drain Height | 650 | mm |
| Required Features of Drainage System | Venting |
Electrical (6)
| Specification | Value | Unit |
|---|---|---|
| Max Frequency | 60 | Hz |
| Max Voltage | 440 | V |
| Max Wattage | 28 | kW |
| Power Of Steam Source | TBC | kW |
| Standard Amperage | 29 | A |
| Three Phase Availability | Available |
Process (2)
| Specification | Value | Unit |
|---|---|---|
| Fitted Vacuum System | Optional | |
| Steam Source | In-chamber water heating element |
Water (3)
| Specification | Value | Unit |
|---|---|---|
| Maximum Water Flow Rate | Infinity | Litres per minute |
| Minimum Water Flow Rate | 0 | Litres per minute |
| Water Enters Via | a water jug, filled from a tap |
Options
Options (30)
| Code | Name | Description | Functionality / Features |
|---|---|---|---|
| RSSC70 | Secondary Controller | A secondary screen used on double-door pass-through autoclaves, allowing the operator to view the controller displayed information from the opposite door. |
|
| AVC001L | Advanced Vacuum | Provides pre- and post-sterilization vacuum phases, using negative and positive pressure pulsing to remove air from the chamber and load before sterilization, and a drying stage after sterilization. When combined with a Chamber Vessel Heating Jacket, drying performance is enhanced. |
|
| AVC002L | Simple Vacuum | Provides a vacuum phase before and after sterilization to aid with air removal and load drying, reducing time to sterilization. Suitable for use with in-chamber steam generation. |
|
| AVC004 | Air Ballast | Maintains chamber pressure during the cooling of fluid loads, greatly reducing the likelihood of boil-over. Requires suitable compressed air, and works best when paired with Load-Sensed Process Timing and a Cooling Jacket. |
|
| AVC007 | Special Pathogen Free Seal | A mechanical seal used in double-door (pass-through) autoclaves, uniting the autoclave with a surrounding wall to maintain separation between the clean and unclean sides of a laboratory. Prevents cross-contamination, allowing the autoclave to function as a pathogen-controlled access point between two locations. |
|
| AVQ006L | CAT III (CL3 / BSL3) Compliance | Enhanced containment methods prevent unsterilized pathogens from leaving the autoclave, meeting the requirements of CAT III compliance for Containment Level 3 / Biosafety Level 3. |
|
| CFR021 | FDA 21 CFR Part 11 Compliant Controller Software | Provides compliance to the requirements of 21 CFR Part 11 through data security, data integrity, traceability/audit trails, and electronic signature rules. |
|
| 24KWSG | 24kW Integral Steam Generator | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| AAN420 | Drain Cooling | Cools effluent (waste water) before it leaves the autoclave. For sites without thermal-shock-resistant drains. Requires a mains water supply. |
|
| AAP006 | Chamber Vessel Cooling Fan System | Blows air across the external surfaces of the sterilization chamber vessel to improve cooling times. |
|
| AAP018 | Autodrain | Automatically drains the in-chamber steam generation assembly at the end of a cycle, allowing the sterilization chamber to cool faster. |
|
| AAP019 | Autofill | Allows automatic refilling of the in-chamber steam generation assembly. Requires mains water supply. |
|
| AAP100 | Chamber Vessel Cooling Water Coils | Pipework coiled around the sterilization chamber vessel allows water to be circulated around the outside, improving cooling times. |
|
| AAP102S | In Chamber Fan | Circulates gases inside the chamber during cooling, allowing the load to cool extremely quickly. Requires a Chamber Vessel Cooling Jacket for optimal performance, and Air Ballast is required when sterilizing liquids. Also required for over-pressure (Steam-Air mix) applications. |
|
| AAQ503 | Integral Air Compressor | Provides compressed air for autoclave systems that require it. | |
| AAR014 | Load Sensed Process Timing | A temperature probe with an armoured lead is added to the chamber. Placed inside fluid loads, this probe provides accurate measurement of liquid temperatures, allowing for faster and more controlled liquid sterilization. |
|
| AAR122 | Ethernet Interface | Allows the autoclave controller to be monitored from compatible digital systems. |
|
| AAR130 | Integral Data-Printer | Allows for the generation of physical records. |
|
| AJP152 | Chamber Vessel Heating Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with steam to heat the chamber walls. Provides improved heating speeds, and load drying after sterilization. |
|
| AJP152C | Chamber Vessel Cooling Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with water to cool the chamber walls. Provides rapid cooling. |
|
| AAQ525 | External Air Compressor | Provides compressed air for autoclave systems that require it. |
|
| AAB001 | Blow Down Vessel | For use with an integral or stand-alone steam generator. The autoclave can be set to expel high-pressure steam to remove dissolved gases and debris from its components; this vessel safely captures the expelled condensate and debris during the blow-down process. |
|
| CWS300 | Water Softener and Installation Kit | Extracts limescale-causing minerals from the water supply before they enter the autoclave, reducing the likelihood of limescale fouling the autoclave. |
|
| AAN080 | Container Tray | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN300 | Morrison Discard Container | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN318 | Additional Shelf | An additional shelf on anti-tilt brackets, typically placed in the centre of the autoclave, providing approximately twice the usable width and depth of chamber space. |
|
| IQ/OQ001 | IQ/OQ Documents Package | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| IQ/OQ003 | Completion of IQ/OQ Documentation | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| UKAS001 | UKAS Calibration | Calibration of the autoclave to specified temperatures as per the requirements of ISO 17025:2005. |
|
| UKASVAL | Load Validation | Multipoint Load Validation Service provided by an Astell Services Autoclave Engineer. |
|
ASB290 - 290 Litre Front Loading Autoclave
Specifications
Air (7)
| Specification | Value | Unit |
|---|---|---|
| Connecting Airline Hose Size For Air Ballast | TBC | mm |
| Peak Air Flow For Air Ballast | 100 | Litres per minute |
| Peak Air Pressure For Air Ballast | TBC | bar |
| Second Air Supply Use | simple or advanced vacuum | |
| Second Airline Supply Connection Size | TBC | mm |
| Second Peak Air Supply Flow | TBC | Litres per minute |
| Second Peak Air Supply Pressure | TBC | bar |
Chamber (5)
| Specification | Value | Unit |
|---|---|---|
| Chamber Depth | 1017 | mm |
| Chamber Finish | Bead Blasted | |
| Chamber Height | 600 | mm |
| Chamber Volume | 290 | Litres |
| Chamber Width | 600 | mm |
Dimensions (5)
| Specification | Value | Unit |
|---|---|---|
| Dry Weight | 392.5 | kg |
| Product Depth | 1440 | mm |
| Product Footprint | 1.3 | square metres |
| Product Height | 1420 | mm |
| Product Width | 902 | mm |
Doors (1)
| Specification | Value | Unit |
|---|---|---|
| Number Of Doors | 1 |
Drain (6)
| Specification | Value | Unit |
|---|---|---|
| Drain Material | Copper | |
| Drain Position | above floor | |
| Drain Size | 35 | mm |
| Drainpipe Connection | Between Autoclave and Main Building Drain Only | |
| Maximum Drain Height | 650 | mm |
| Required Features of Drainage System | Venting |
Electrical (6)
| Specification | Value | Unit |
|---|---|---|
| Max Frequency | 60 | Hz |
| Max Voltage | 440 | V |
| Max Wattage | 16 | kW |
| Power Of Steam Source | TBC | kW |
| Standard Amperage | 38 | A |
| Three Phase Availability | Available |
Process (1)
| Specification | Value | Unit |
|---|---|---|
| Steam Source | In-chamber water heating element |
Water (3)
| Specification | Value | Unit |
|---|---|---|
| Maximum Water Flow Rate | Infinity | Litres per minute |
| Minimum Water Flow Rate | 0 | Litres per minute |
| Water Enters Via | a water jug, filled from a tap |
Options
Options (30)
| Code | Name | Description | Functionality / Features |
|---|---|---|---|
| AAR122 | Ethernet Interface | Allows the autoclave controller to be monitored from compatible digital systems. |
|
| AAR130 | Integral Data-Printer | Allows for the generation of physical records. |
|
| AJP152 | Chamber Vessel Heating Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with steam to heat the chamber walls. Provides improved heating speeds, and load drying after sterilization. |
|
| AJP152C | Chamber Vessel Cooling Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with water to cool the chamber walls. Provides rapid cooling. |
|
| AVC001L | Advanced Vacuum | Provides pre- and post-sterilization vacuum phases, using negative and positive pressure pulsing to remove air from the chamber and load before sterilization, and a drying stage after sterilization. When combined with a Chamber Vessel Heating Jacket, drying performance is enhanced. |
|
| AVC002L | Simple Vacuum | Provides a vacuum phase before and after sterilization to aid with air removal and load drying, reducing time to sterilization. Suitable for use with in-chamber steam generation. |
|
| AVC004 | Air Ballast | Maintains chamber pressure during the cooling of fluid loads, greatly reducing the likelihood of boil-over. Requires suitable compressed air, and works best when paired with Load-Sensed Process Timing and a Cooling Jacket. |
|
| AVC007 | Special Pathogen Free Seal | A mechanical seal used in double-door (pass-through) autoclaves, uniting the autoclave with a surrounding wall to maintain separation between the clean and unclean sides of a laboratory. Prevents cross-contamination, allowing the autoclave to function as a pathogen-controlled access point between two locations. |
|
| AVQ006L | CAT III (CL3 / BSL3) Compliance | Enhanced containment methods prevent unsterilized pathogens from leaving the autoclave, meeting the requirements of CAT III compliance for Containment Level 3 / Biosafety Level 3. |
|
| CFR021 | FDA 21 CFR Part 11 Compliant Controller Software | Provides compliance to the requirements of 21 CFR Part 11 through data security, data integrity, traceability/audit trails, and electronic signature rules. |
|
| AAN420 | Drain Cooling | Cools effluent (waste water) before it leaves the autoclave. For sites without thermal-shock-resistant drains. Requires a mains water supply. |
|
| AAP006 | Chamber Vessel Cooling Fan System | Blows air across the external surfaces of the sterilization chamber vessel to improve cooling times. |
|
| AAP018 | Autodrain | Automatically drains the in-chamber steam generation assembly at the end of a cycle, allowing the sterilization chamber to cool faster. |
|
| AAP019 | Autofill | Allows automatic refilling of the in-chamber steam generation assembly. Requires mains water supply. |
|
| AAP100 | Chamber Vessel Cooling Water Coils | Pipework coiled around the sterilization chamber vessel allows water to be circulated around the outside, improving cooling times. |
|
| AAP102S | In Chamber Fan | Circulates gases inside the chamber during cooling, allowing the load to cool extremely quickly. Requires a Chamber Vessel Cooling Jacket for optimal performance, and Air Ballast is required when sterilizing liquids. Also required for over-pressure (Steam-Air mix) applications. |
|
| AAQ503 | Integral Air Compressor | Provides compressed air for autoclave systems that require it. | |
| 24KWSG | 24kW Integral Steam Generator | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| RSSC70 | Secondary Controller | A secondary screen used on double-door pass-through autoclaves, allowing the operator to view the controller displayed information from the opposite door. |
|
| AAR014 | Load Sensed Process Timing | A temperature probe with an armoured lead is added to the chamber. Placed inside fluid loads, this probe provides accurate measurement of liquid temperatures, allowing for faster and more controlled liquid sterilization. |
|
| AAB001 | Blow Down Vessel | For use with an integral or stand-alone steam generator. The autoclave can be set to expel high-pressure steam to remove dissolved gases and debris from its components; this vessel safely captures the expelled condensate and debris during the blow-down process. |
|
| AAQ525 | External Air Compressor | Provides compressed air for autoclave systems that require it. |
|
| CWS300 | Water Softener and Installation Kit | Extracts limescale-causing minerals from the water supply before they enter the autoclave, reducing the likelihood of limescale fouling the autoclave. |
|
| AAN080 | Container Tray | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN300 | Morrison Discard Container | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN318 | Additional Shelf | An additional shelf on anti-tilt brackets, typically placed in the centre of the autoclave, providing approximately twice the usable width and depth of chamber space. |
|
| UKAS001 | UKAS Calibration | Calibration of the autoclave to specified temperatures as per the requirements of ISO 17025:2005. |
|
| IQ/OQ001 | IQ/OQ Documents Package | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| IQ/OQ003 | Completion of IQ/OQ Documentation | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| UKASVAL | Load Validation | Multipoint Load Validation Service provided by an Astell Services Autoclave Engineer. |
|
ASB300 - 344 Litre Front Loading Autoclave
Specifications
Air (7)
| Specification | Value | Unit |
|---|---|---|
| Connecting Airline Hose Size For Air Ballast | TBC | mm |
| Peak Air Flow For Air Ballast | 100 | Litres per minute |
| Peak Air Pressure For Air Ballast | TBC | bar |
| Second Air Supply Use | simple or advanced vacuum | |
| Second Airline Supply Connection Size | TBC | mm |
| Second Peak Air Supply Flow | TBC | Litres per minute |
| Second Peak Air Supply Pressure | TBC | bar |
Chamber (5)
| Specification | Value | Unit |
|---|---|---|
| Chamber Depth | 1217 | mm |
| Chamber Finish | Bead Blasted | |
| Chamber Height | 600 | mm |
| Chamber Volume | 344 | Litres |
| Chamber Width | 600 | mm |
Dimensions (5)
| Specification | Value | Unit |
|---|---|---|
| Dry Weight | 418 | kg |
| Product Depth | 1630 | mm |
| Product Footprint | 1.47 | square metres |
| Product Height | 1420 | mm |
| Product Width | 902 | mm |
Doors (1)
| Specification | Value | Unit |
|---|---|---|
| Number Of Doors | 1 |
Drain (6)
| Specification | Value | Unit |
|---|---|---|
| Drain Material | Copper | |
| Drain Position | above floor | |
| Drain Size | 35 | mm |
| Drainpipe Connection | Between Autoclave and Main Building Drain Only | |
| Maximum Drain Height | 650 | mm |
| Required Features of Drainage System | Venting |
Electrical (6)
| Specification | Value | Unit |
|---|---|---|
| Max Frequency | 60 | Hz |
| Max Voltage | 440 | V |
| Max Wattage | 16 | kW |
| Power Of Steam Source | TBC | kW |
| Standard Amperage | 35 | A |
| Three Phase Availability | Available |
Process (2)
| Specification | Value | Unit |
|---|---|---|
| Fitted Vacuum System | Optional | |
| Steam Source | In-chamber water heating element |
Water (3)
| Specification | Value | Unit |
|---|---|---|
| Maximum Water Flow Rate | Infinity | Litres per minute |
| Minimum Water Flow Rate | 0 | Litres per minute |
| Water Enters Via | a water jug, filled from a tap |
Options
Options (30)
| Code | Name | Description | Functionality / Features |
|---|---|---|---|
| AAR130 | Integral Data-Printer | Allows for the generation of physical records. |
|
| AAN420 | Drain Cooling | Cools effluent (waste water) before it leaves the autoclave. For sites without thermal-shock-resistant drains. Requires a mains water supply. |
|
| AAP006 | Chamber Vessel Cooling Fan System | Blows air across the external surfaces of the sterilization chamber vessel to improve cooling times. |
|
| AAP018 | Autodrain | Automatically drains the in-chamber steam generation assembly at the end of a cycle, allowing the sterilization chamber to cool faster. |
|
| AAP019 | Autofill | Allows automatic refilling of the in-chamber steam generation assembly. Requires mains water supply. |
|
| AAP100 | Chamber Vessel Cooling Water Coils | Pipework coiled around the sterilization chamber vessel allows water to be circulated around the outside, improving cooling times. |
|
| AAP102S | In Chamber Fan | Circulates gases inside the chamber during cooling, allowing the load to cool extremely quickly. Requires a Chamber Vessel Cooling Jacket for optimal performance, and Air Ballast is required when sterilizing liquids. Also required for over-pressure (Steam-Air mix) applications. |
|
| AAQ503 | Integral Air Compressor | Provides compressed air for autoclave systems that require it. | |
| AAR014 | Load Sensed Process Timing | A temperature probe with an armoured lead is added to the chamber. Placed inside fluid loads, this probe provides accurate measurement of liquid temperatures, allowing for faster and more controlled liquid sterilization. |
|
| AAR122 | Ethernet Interface | Allows the autoclave controller to be monitored from compatible digital systems. |
|
| AJP152 | Chamber Vessel Heating Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with steam to heat the chamber walls. Provides improved heating speeds, and load drying after sterilization. |
|
| AJP152C | Chamber Vessel Cooling Jacket | The sterilization chamber vessel is encapsulated within another vessel, and the space between can be filled with water to cool the chamber walls. Provides rapid cooling. |
|
| 24KWSG | 24kW Integral Steam Generator | Replaces the in-chamber steam generation assembly with hardware positioned outside the sterilization chamber to provide enhanced steam control and functionality. Requires mains water supply. |
|
| AVC001L | Advanced Vacuum | Provides pre- and post-sterilization vacuum phases, using negative and positive pressure pulsing to remove air from the chamber and load before sterilization, and a drying stage after sterilization. When combined with a Chamber Vessel Heating Jacket, drying performance is enhanced. |
|
| AVC002L | Simple Vacuum | Provides a vacuum phase before and after sterilization to aid with air removal and load drying, reducing time to sterilization. Suitable for use with in-chamber steam generation. |
|
| AVC004 | Air Ballast | Maintains chamber pressure during the cooling of fluid loads, greatly reducing the likelihood of boil-over. Requires suitable compressed air, and works best when paired with Load-Sensed Process Timing and a Cooling Jacket. |
|
| AVC007 | Special Pathogen Free Seal | A mechanical seal used in double-door (pass-through) autoclaves, uniting the autoclave with a surrounding wall to maintain separation between the clean and unclean sides of a laboratory. Prevents cross-contamination, allowing the autoclave to function as a pathogen-controlled access point between two locations. |
|
| AVQ006L | CAT III (CL3 / BSL3) Compliance | Enhanced containment methods prevent unsterilized pathogens from leaving the autoclave, meeting the requirements of CAT III compliance for Containment Level 3 / Biosafety Level 3. |
|
| CFR021 | FDA 21 CFR Part 11 Compliant Controller Software | Provides compliance to the requirements of 21 CFR Part 11 through data security, data integrity, traceability/audit trails, and electronic signature rules. |
|
| RSSC70 | Secondary Controller | A secondary screen used on double-door pass-through autoclaves, allowing the operator to view the controller displayed information from the opposite door. |
|
| CWS300 | Water Softener and Installation Kit | Extracts limescale-causing minerals from the water supply before they enter the autoclave, reducing the likelihood of limescale fouling the autoclave. |
|
| AAB001 | Blow Down Vessel | For use with an integral or stand-alone steam generator. The autoclave can be set to expel high-pressure steam to remove dissolved gases and debris from its components; this vessel safely captures the expelled condensate and debris during the blow-down process. |
|
| AAQ525 | External Air Compressor | Provides compressed air for autoclave systems that require it. |
|
| AAN318 | Additional Shelf | An additional shelf on anti-tilt brackets, typically placed in the centre of the autoclave, providing approximately twice the usable width and depth of chamber space. |
|
| AAN080 | Container Tray | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| AAN300 | Morrison Discard Container | A cuboid container that allows the loading and unloading of multiple items simultaneously. |
|
| UKAS001 | UKAS Calibration | Calibration of the autoclave to specified temperatures as per the requirements of ISO 17025:2005. |
|
| IQ/OQ003 | Completion of IQ/OQ Documentation | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| IQ/OQ001 | IQ/OQ Documents Package | Completion of the Installation Qualification (IQ) and Operational Qualification (OQ) documentation by an Astell Service Engineer. Requires verification to be valid. |
|
| UKASVAL | Load Validation | Multipoint Load Validation Service provided by an Astell Services Autoclave Engineer. |
|